[Home ] [Archive]    
:: Main :: About :: Current Issue :: Archive :: Search :: Submit :: Contact ::
Main Menu
Home::
IJRR Information::
For Authors::
For Reviewers::
Subscription::
News & Events::
Web Mail::
::
Search in website

Advanced Search
..
Receive site information
Enter your Email in the following box to receive the site news and information.
..
ISSN
Hard Copy 2322-3243
Online 2345-4229
..
Online Submission
Now you can send your articles to IJRR office using the article submission system.
..

AWT IMAGE

AWT IMAGE

:: Volume 12, Issue 3 (7-2014) ::
Int J Radiat Res 2014, 12(3): 249-256 Back to browse issues page
Monte Carlo study on beam hardening effect of physical wedges
Gh. Geraily , M. Mirzapour , S.R. Mahdavi , M. Allahverdi , A. Mostaar , M. Masoudifar
Department of Medical Physics, Tehran University of Medical Science, Tehran, Iran , srmahdavi@hotmail.com
Abstract:   (8147 Views)

Background: Physical wedges are still widely used as beam modifiers in external beam radiotherapy. However the presence of them in the beam trace may cause beam hardening which may not be considered in many treatment planning systems. The aim of this study is to investigate the beam hardening effect generated by physical wedges via different beam quality indexes as photon spectrum, half value layer, mean energy and tissue–phantom ratio. Materials and Methods: The effect of physical wedges on the photon beam quality of a 6-18MV Varian 2100C/D accelerator was studied with the BEAMnrc Monte Carlo code. Good agreements were obtained between measured and calculated depth doses and beam profiles for open and wedged photon beams at both energies. Results: It was noticed that for 6 MV photon beams, physical wedges have more significant effects on beam quality than for 18 MV. Also it was obtained that at 18 MV photon beam as the wedge angle increased, the effect of wedge on beam quality becomes reversed and beam softening occurred. Conclusion: According to these results, it is recommended that beam hardening and softening of physical wedges should be considered in treatment planning systems in order to increase the accuracy in dose delivery.

Keywords: Monte Carlo, beam hardening, physical wedge
Full-Text [PDF 574 kb]   (2324 Downloads)    
Type of Study: Original Research | Subject: Medical Physics
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA


XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Geraily G, Mirzapour M, Mahdavi S, Allahverdi M, Mostaar A, Masoudifar M. Monte Carlo study on beam hardening effect of physical wedges. Int J Radiat Res 2014; 12 (3) :249-256
URL: http://ijrr.com/article-1-1283-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 12, Issue 3 (7-2014) Back to browse issues page
International Journal of Radiation Research
Persian site map - English site map - Created in 0.09 seconds with 50 queries by YEKTAWEB 4645