:: Volume 3, Issue 4 (March 2006) ::
Int J Radiat Res 2006, 3(4): 153-161 Back to browse issues page
Achievable accuracy in brain tumors by in vivo dosimetry with diode detectors
M. Allahverdi , M.R. Taghizadeh
, malahverdi@yahoo.co.uk
Abstract:   (16272 Views)

ABSTRACT

Background: In vivo measurements of applied dose during radiotherapy treatment, is important to ensure accurate dose delivery to patients. Uncertainty in dose delivery should fall within ±5% of the prescribed dose as recommended by ICRU. Assessment of dose for radiotherapy applications performed with various types of detectors. In this study, semiconductor diodes were used which have some advantages for clinical dosimetry. Materials and Methods: The brain tumors have generally treated with two fields using SSD technique. Entrance and exit dose were measured for each patient with diodes during treatment. Entrance and exit dose measurements have converted to midline dose. Measured entrance and exit doses have compared with calculated ones and large deviations (more than 5%) have observed. A farmer ionization chamber (0.6 cm) was used as the reference dose detector and a Perspex water phantom (30 30cm2 area and thickness ranging from 5 cm to 30 cm) were used to determine calibration and correction factors. Results: Correction factors were determined and variations more than 1% have used to obtain correct doses. Large deviations between measured and calculated for entrance (5.3%), exit (42%) and midline (47%) were detected. The difference was not found to be significant when comparing the measured entrance dose with the calculated one (p=0.696) and the measured exit dose with calculated one (p=0.643) and measured midline dose with calculated one (p=0.104). Conclusion: In vivo dosimetry is very useful to check the dose delivered to the patient. A high precision obtained when the calibration and correction factors for each parameter of influence on the diode response are carefully determined and applied to convert the diode signal in the adsorbed dose. In this study, the target-absorbed doses were estimated from the measured entrance dose and the measured transmission.

Keywords: In vivo dosimetry, brain tumors, semiconductor detectors
Full-Text [PDF 415 kb]   (2568 Downloads)    
Type of Study: Original Research |
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA


XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

M. Allahverdi, M.R. Taghizadeh. Achievable accuracy in brain tumors by in vivo dosimetry with diode detectors. Int J Radiat Res 2006; 3 (4) :153-161
URL: http://ijrr.com/article-1-187-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 3, Issue 4 (March 2006) Back to browse issues page
International Journal of Radiation Research
Persian site map - English site map - Created in 0.06 seconds with 49 queries by YEKTAWEB 4645