TY - JOUR T1 - Enhancement of MD-55-2 radiochromic film sensitivity using a multilayer film technique for applications in the low dose range TT - JF - Int-J-Radiat-Res JO - Int-J-Radiat-Res VL - 3 IS - 1 UR - http://ijrr.com/article-1-117-en.html Y1 - 2005 SP - 11 EP - 16 N2 -  ABSTRACTBackgrounds: MD-55-2 is one of the Radiochromic film models with the sensitivity suitable for dose measurements ranging from 5 to 100 Gy.  However, this lower limit makes the film impractical for its applications in many areas such as brachytherapy source dosimetry.Materials and methods: In this project, the useful range of the film has been extended by using a multilayer film technique. In this technique, single-, double-, and triple- layers of films were exposed to the doses ranging from 0.5 to 10 Gy using a Co-60 photon beam. Calibration curves for corresponding layers of films were obtained with a spectrophotometer using a 680nm wavelength.Results: The results indicated that the sensitivities of double and triple layers were approximately 200% and 300%, respectively, higher than a single-layer film.  The impact of multilayer film arrangement on the energy dependence of the MD-55-2 Radiochromic film has also been examined using 100KVp, 80 KVp, and 6 MV X-ray beams. The results indicated an insignificant (within 5%) change in film responses with the beam energy. Conclusion: Therefore, the multilayer technique enhances the Radiochromic film sensitivity and expands its application to the low dose range in field of brachytherapy source dosimetry.    M3 ER -