Feasibility study of multi-purpose quality assurance phantom for pretreatment verification of volumetric modulated arc therapy
H.S. Won , J.B. Chung , K.Y. Eom , D.G. Hwang , S.W. Kang , T. S. Suh
Department of Radiation Oncology, Seoul National University Bungdang Hospital, Seongnam, Korea , jbchung1213@gmail.com
Abstract:   (5022 Views)
Background: The purpose of this study was to evaluate the feasibility of a multi-purpose quality assurance (QA) phantom for pretreatment verification of volumetric modulated arc therapy (VMAT). Materials and Methods: The QA phantom was constructed with polymethyl methacrylate (PMMA) to perform relative dosimetry using EBT3 film and MapCHECK, as well as absolute dosimetry using an ionization chamber. The QA phantom was constructed to perform relative dosimetry using EBT3 film and MapCHECK, as well as the absolute dosimetry using ionization chamber. In order to verify the pretreatment plans, 25 patients treated with VMAT were selected. The pretreatment plans were calculated in the Eclipse treatment planning system using the Acuros XB dose calculation algorithm and CT images for the QA phantom, with the same beam setup and monitor units (MUs) as those for patient treatment. All plans were delivered to the Varian TrueBeam accelerator equipped with a high-definition multi-leaf collimator. Results: The multi-purpose QA phantom is developed for convenient VMAT dose verification. By using the QA phantom, all 25 cases passed ±3% acceptability criteria in absolute dosimetry with an ionization chamber for pretreatment verification. The relative dosimetry using EBT3 film and MapCHECK system also showed high agreement of more than 90% for 2%/2-mm and 3%/3-mm criteria. Conclusion: The results of this study demonstrated the good multi-purpose capabilities of the phantom for the absolute and relative dosimetry. Therefore, the developed multi-purpose QA phantom was applied in our institution for routine VMRT dose verification.
 
Keywords: Volumetric modulated arc therapy, phantom, absolute dosimetry, relative dosimetry.
Full-Text [PDF 1989 kb]   (1115 Downloads)    
Type of Study: Original Research | Subject: Radiation Biology


XML     Print



Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 16, Issue 3 (7-2018) Back to browse issues page