Designing a shield to reduce radiation dose during mammography: Dosimetric evaluation
Radiation Biology
Radiation Biology
تحقيق بديع
Original Research
<div style="text-align: justify;">Background: This paper presents a method to reduce radiation exposure during mammography by analysing the doses to ipsilateral and contralateral breasts and to adjacent organs by evaluating material-dependent shielding performance. Materials and Methods: Six target-filter combinations (Mo-Mo, Mo-Rh, Rh-Rh, Rh-Mo, W-Mo, W-Rh) were tested by measuring the doses delivered to the breasts and adjacent organs, with the contralateral (opposite side) breast shielded. The shield was designed to have a simple (┓,┎) shape for ease of use in actual clinical settings, using lead, copper, bismuth, and barium sulphate (BaSO4) as materials for shield configuration. Results: The dosimetric data revealed that the highest absorbed dose was exhibited by the target filter combination of Rh/Rh, followed by W/Rh, W/Mo, Rh/Mo, Mo/Rh, and Mo/Mo. Additionally, the radiation dose was reduced by 54–55%, with the average absorbed dose on the contralateral breast reduced from 0.655 to 0.359 mGy. All four shielding materials used in the experiments were analysed for the shielding effect. Conclusion: This Using a shield during screening mammography would alleviate concerns about the mammography-induced risk of breast cancer and secondary effects. </div>
Mammography, Simulation, Shielding material, Contralateral dose
127
135
http://ijrr.com/browse.php?a_code=A-10-1-765&slc_lang=en&sid=1
D.Y.
Lee
7900319475328460013220
7900319475328460013220
No
Department of Radiation Oncology, Dongnam Institute of Radiological & Medical Sciences Cancer Center, 40, Jwadong-gil, Busan, Republic of Korea
J.S.
Lee
sonojinsoo@naver.com
7900319475328460013221
7900319475328460013221
Yes
Department of Radiology, Inje University Haeundae Paik Hospital, 875, Haeundae-ro, Haeundae-gu, Busan, Republic of Korea