Int J Radiat Res 2026, 24(2): 373-379 Back to browse issues page
Effect of a flattening filter on the surface and build-up doses for high-energy photons produced with a varian TrueBeam STx linear accelerator
F. Alkaya , C.K. Akbas , S.Ç. Karacam , H.B. Becerir
Department of Healt Physics,Graduate School of Health Sciences, Istanbul Medipol University, 34810, Istanbul, Turkey
Abstract:   (599 Views)
Background: The aim of thise study was to measure and determine the surface  and build- up region doses at different field sizes on the central axis of 6 and 10 MV with flettening filter (FF) and without flettenin filter (FFF) high-energy photon beams, which are commonly used in clinical radiotherapy applications, and to reveal the difference between FF  and FFF beams. Material and Methods: The dose measurements were made using a parallel-plate ion chamber and a water-equivalent solid water phantom for a field size of 5 × 5, 10 × 10, and 20 × 20 cm2, with a source skin distance of 100 cm. Results: The relative surface doses for 6 MV FF high-energy photons are 10.9%±0.007%, 16.4%±0.010% and 26.9%±0.014% for field sizes of 5x5, 10x10 and 20x20 cm2. For 6 MV FFF they are 14.7%±0.010%, 19.7%±0.010% and 27.2±0.009% respectively. The relative surface doses for 10 MV FF high-energy photons are 6.3%±0.010%, 12.0%±0.007% and 23.7%±0.010% for field sizes of 5x5, 10x10 and 20x20 cm2. For 10 MV FFF they are 9.7%±0.010%, 13.8%±0.007% and 20.1%±0.007% respectively. Conclusion: It was observed that the relative surface doses increased in FFF for 6MV compared to FF photon beams, and increased in 5x5 and 10x10 cm2 field sizes and decreased in 20x20 cm2 field size for 10MV photon beam. These results may have important implications in radiotherapy applications where skin dose is important.
Keywords: TrueBeam STx, flattening filter free, surface dose, build-up dose, radiotherapy.
Full-Text [PDF 823 kb]   (216 Downloads)    
Subject: Radiation Biology
References
1. Delaney GP, Jacob S, Featherstone C, Barton MB, (2005) The role of radiotherapy in Cancer care: estimating optimal utilization from a review of evidence based clinical guide lines. Cancer, 104(6): 1129-37. [DOI:10.1002/cncr.21324]
2. Kragl G, Wetterstedt SA, Knäusl B, et al. (2009) Dosimetric characteristics of 6 and 10 MV unflattened photon beams. Radiother Oncol, 93(1): 141-146. [DOI:10.1016/j.radonc.2009.06.008]
3. Cashmore J (2016) Surface dose variations in 6 and 10 MV flattened and flattening filter-free (FFF) photon beams. J Appl Clin Med Phys, 17(5): 293-307. [DOI:10.1120/jacmp.v17i5.6284]
4. Poffenbarger B (2010) Evaluation of the new Varian Trilogy MX flattening filter free 6 MV and 10 MV photon beams. Med Phys, 37(6-26): Abstract ID: 14162. [DOI:10.1118/1.3469189]
5. Balik S, Qi P, Magnelli A, et al. (2018) Comparison of spine SRS VMAT plans with flattening filter free 6 MV or 10 MV beams. Int J Radiat Oncol Biol Phys, 102: (3S Supplement). [DOI:10.1016/j.ijrobp.2018.07.1436]
6. Lang S, Shrestha B, Graydon S (2013) Clinical application of flattening filter free beams for extracranial stereotactic radiotherapy. Radiother Oncol, 106(2): 255-259. [DOI:10.1016/j.radonc.2012.12.012]
7. Whitten G, Daly U, McCallum CD, et al. (2022) Flattening filter free stereotactic radiosurgery for brain metastases using dynamic conformal arcs: 6 MV or 10 MV?. J Radiother Pract, 21(3): 328-334. [DOI:10.1017/S1460396920001247]
8. Lamb A, Blake S, (1998) Investigation and modelling of the surface dose from linear accelerator produced 6 and 10 MV photon beams. Phys Med Biol, 43(5): 1133-1146. [DOI:10.1088/0031-9155/43/5/006]
9. Braunstein S and Nakamura JL ( 2013) Radiotherapy-induced malignancies: review of clinical features, pathobiology, and evolving approaches for mitigating risk. Frontiers in Oncology/Radiation Oncology, 3(73): 1-15. [DOI:10.3389/fonc.2013.00073]
10. Akbas U, Kesen N.D, Koksal C, et al. (2016) Surface and build-up region dose measurements with Markus parallel-plate ionization chamber GafCromic EBT3 film and MOSFET detectorr for high energy photon beams. Advances in High Energy Physics, Article ID 8361028-10 pages. [DOI:10.1155/2016/8361028]
11. Gerbi BJ and Khan FM (1990) Measurement of dose in the build-up region using fixed-separation parallel-plane ionization chambers. Med Phys, 17(1): 17-26. [DOI:10.1118/1.596522]
12. Imae T, Takenaka S, Watanabe Y, et al. (2020) Surface and build‐up dose comparison between Elekta 6 MV flattening filter and flattening‐filter‐free beams using an advanced Markus ionization chamber and a solid water‐equivalent phantom. J Appl Clin Med Phys, 21(12): 334-339. [DOI:10.1002/acm2.13094]
13. Guy CL, Karki K, Sharma M, Kim S (2016) Clinically relevant investigation of flattening filter-free skin dose. J Appl Clin Med Phys, 17(6): 140-148. [DOI:10.1120/jacmp.v17i6.6307]
14. Sigamani A, Nambiraj A,Yadav G, et al. (2016) Surface dose measurements and comparison of unflattened and flattened photon beams. J Med Phys, 41(2): 85-91. [DOI:10.4103/0971-6203.181648]
15. Yani S, Budiansah I, Pratama SH, et al. (2021) Evaluation of the dosimetric characteristics of 10 MV flattened and unflattened photon beams in a heterogeneous phantom. Int J Radiat Res, 19(4): 835-841. [DOI:10.52547/ijrr.19.4.9]
16. Abdel-Rahman W, Seuntjens JP, Verhaegen F, et al. (2005) Validation of Monte Carlo calculated surface doses for megavoltage photon beams. Med Phys, 32(1): 286-298. [DOI:10.1118/1.1829401]
17. Bilge H, Cakir A, Okutan M, Acar H (2009) Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams. Phys Med, 25(2): 101-104. [DOI:10.1016/j.ejmp.2008.05.001]
18. Butson MJ, Cheung T, Yu PKN, Currie M (2004) Surface dose extrapolation measurements with radiographic film. Phys Med Biol, 49(13): N197-N201. [DOI:10.1088/0031-9155/49/13/N03]
19. Kesen ND and Koksal C (2020) Investigation of surface and build-up region doses for 6 MV high energy photon beams in the presence of a thermoplastic mask. Int J Radiat Res, 18(4): 623-631. [DOI:10.52547/ijrr.18.4.623]
20. Kim JH, Hill R, Kuncic Z (2012) Practical considerations for reporting surface dose in external beam radiotherapy: a 6 MV X-ray beam study. Australas Phys Eng Sci Med, 35(3): 271-282. [DOI:10.1007/s13246-012-0145-1]
21. Gerbi BJ and Khan FM (1987) The polarity effect for commercially available parallel-plane ionization chamber. Med Phys, 14(2): 210-5. [DOI:10.1118/1.596072]
22. Nambiraj A, Sigamani A, Kaviarasu K, et al. (2015) Surface dose and build-up dose comparison of 6MV and 10MV Flattened with 6MV FFF and 10MV FFF photon beams with different detectors. Med Phys, 42 (6-23): 3502-3502. [DOI:10.1118/1.4925082]
23. Richmond N, Allen V, Daniel J, et al. (2015) A comparison of phantom scatter from flattened and flattening filter free high-energy photon beams. Med Dosim, 40(1): 58-63. [DOI:10.1016/j.meddos.2014.10.001]
24. Butson MJ and Yu PKN (1999) Meltcalfe EM. Extrapolated surface dose measurements with radiochromic film. Med Phys, 26: 485-7. [DOI:10.1118/1.598539]
25. Devic S, Seuntjens J, Abdel-Rahman W, et al. (2006) Accurate skin dose measurements using radiochromic film in clinical applications. Med Phys, 33(4): 1116-23. [DOI:10.1118/1.2179169]
26. Apipunyasopan L, Srisatit S, Phaisangittisakul N (2013) An investigation of the depth dose in the build-up region, and surface dose for a 6-MV therapeutic photon beam: Monte Carlo simulation and measurements. J Radiat Res, 54(2): 374-382. [DOI:10.1093/jrr/rrs097]
27. Mohammed Amin SS, Faraj KA (2023) Estimation of oesophageal surface dose in breast cancer patients undergoing supraclavicular irradiation by thermoluminescent dosimeter (TLD) and treatment planning system (TPS). Int J Radiat Res, 21(4): 647-652. [DOI:10.61186/ijrr.21.4.647]
28. Saadatmand P, Mahdavil SR, Nikoofar AG, et al. (2025) Physical aspects of skin dose distribution in tomotherapy of breast cancer. Int J Radiat Res, 23(1): 83-90. [DOI:10.61186/ijrr.23.1.83]
29. Afereydoon S, Enferadi-Aliabad MJE, Mahdie Gholizadeh M, et al. (2025) Comparative surface dosimetry in breast cancer in 3DCRT radiotherapy:TLD measurements and TPS evaluation across techniques. Radiation Physics and Chemistry, 229(112559): 1-9. [DOI:10.1016/j.radphyschem.2025.112559]
Add your comments about this article
Your username or Email:

CAPTCHA



XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Alkaya F, Akbas C, Karacam S, Becerir H. Effect of a flattening filter on the surface and build-up doses for high-energy photons produced with a varian TrueBeam STx linear accelerator. Int J Radiat Res 2026; 24 (2) :373-379
URL: http://ijrr.com/article-1-7015-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 24, Issue 2 (4-2026) Back to browse issues page
International Journal of Radiation Research
Persian site map - English site map - Created in 0.13 seconds with 49 queries by YEKTAWEB 4741